Mtest:amask:ydelta - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:MTESt:AMASk:YDELta

Command Syntax
:MTESt:AMASk:YDELta <value>
<value> ::= Y delta value in NR3 format
The :MTESt:AMASk:YDELta command sets the vertical tolerance around
the waveform for the automasking feature. The absolute value of the
tolerance will be added and subtracted to vertical values of the waveform
to determine the boundaries of the mask.
The vertical tolerance value is interpreted based on the setting specified
by the :MTESt:AMASk:UNITs command; thus, if you specify 250- E3, the
setting for :MTESt:AMASk:UNITs is CURRent, and the current setting
specifies voltage in the vertical direction, the tolerance will be ±250 mV. If
the setting for :MTESt:AMASk:UNITs is DIVisions, the same Y delta value
will set the tolerance to ±250 millidivisions, or 1/4 of a division.
Query Syntax
:MTESt:AMASk:YDELta?
The :MTESt:AMASk:YDELta? query returns the current setting of the Y
tolerance for automasking. If your computer program will interpret this
value, it should also request the current measurement system using the
:MTESt:AMASk:UNITs query.
Return Format
<value><NL>
<value> ::= Y delta value in NR3 format
See Also
Example Code
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :MTESt Commands"
":MTESt:AMASk:UNITs"
":MTESt:AMASk:XDELta"
":MTESt:AMASk:SOURce"
":MTESt:SOURce"
on page 539
"Example Code"
on page 509
on page 509
on page 515
on page 516
on page 514
23
:MTESt Commands
517

Advertisement

Table of Contents
loading

Table of Contents