:MTESt:RMODe:FACTion:SAVE
Command Syntax
:MTESt:RMODe:FACTion:SAVE <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:SAVE command sets saving on mask failures
on or off.
Setting :MTESt:RMODe:FACTion:SAVE ON automatically sets
N O T E
:MTESt:RMODe:FACTion:PRINt OFF.
See
information on save options.
Query Syntax
:MTESt:RMODe:FACTion:SAVE?
The :MTESt:RMODe:FACTion:SAVE? query returns the current mask failure
save setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also
•
•
•
•
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
Chapter
27, ":SAVE Commands," starting on page 621 for more
"Introduction to :MTESt Commands"
":MTESt:RMODe:FACTion:MEASure"
":MTESt:RMODe:FACTion:PRINt"
":MTESt:RMODe:FACTion:STOP"
on page 509
on page 527
on page 528
on page 530
23
:MTESt Commands
529