:OPEE (Operation Status Enable Register)
Command Syntax
:OPEE <mask>
<mask> ::= 15-bit integer
The :OPEE command sets a mask in the Operation Status Enable register.
Set any of the following bits to "1" to enable bit 7 in the Status Byte
Register and potentially cause an SRQ (Service Request) interrupt to be
generated.
From Hardware
Event Registers
OR
To OPER bit in
Status Byte
Register
Table 77 Operation Status Enable Register (OPEE)
Bit
14
13
12
11
10
9
8-6
5
4
3
2-0
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
From Overload
From Mask Test
Event Registers
Event Registers
IOF
IOC
HWE
OVLR
MTE
14
13
12
11
9
IOF
IOC
HWE
OVLR
MTE
14
13
12
11
10
9
8
+
Name
Description
IOF
IO Operation Failed
IOC
IO Operation Complete
HWE
Hardware Event
OVLR
Overload
---
---
MTE
Mask Test Event
---
---
Wait
Wait Trig
Trig
---
---
Run
Running
---
---
Arm
AER?
Reg
Run bit set if oscilloscope not stopped
Wait
:OPEReration:CONDition?
Run
Trig
Operation Status Condition Register
5
3
Wait
:OPERation[:EVENt]?
Run
Trig
Operation Status Event Register
7
6
5
4
3
2
1
0
:OPEE
:OPEE?
Operation Status Enable
(Mask) Register
When Set (1 = High = True), Enables:
Event when the IO operation fails.
Event when the IO operation completes.
Event when hardware event occurs.
Event when 50 input overload occurs.
(Not used.)
Event when mask test event occurs.
(Not used.)
Event when the trigger is armed.
(Not used.)
Event when the oscilloscope is running (not
stopped).
(Not used.)
6
Root (:) Commands
209