Disturbance Analyzer Meas Result - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

12  Disturbance Analyzer Measurement
Save
Notes
Initial S/W Revision
Disturbance List File
A Disturbance List file contains a copy of the disturbance lists obtained during measurement. The first four
lines are system required header lines, and must be in the correct order.
The data follows as comma separated values. A keyword "DATA" precedes the data. Below is an example
of disturbance list file
DisturbanceList
EMI:DAN
A.11.05,N9020A
B1X P13 EA3 BBA B2X ERC UK6 BAB EMC K03 B25 503 P08 526 508 513 S40 P26 B1C HBA P03 PFR ,01
Y Axis Unit,DecibelMicrovolt
DATA
NO,DURATION,PEAK AMPTD,QP AMPTD,TYPE OF DISTURBANCE,TIMESTAMP,COMMENT
1,10.01,87.88, 78.88,CLICK,2011/11/11 11:11:1,
2,9.99,87.88, 78.88,CLICK,2011/11/11 11:11:1,
3,2089.98,87.88, 78.88,CONT DISTURBANCE,2011/11/11 11:11:1,UNDER INVESTIGATION
4,10.02,87.88, 78.88,CLICK,2011/11/11 11:11:1,
5,210.04,87.88, 78.88,CONT DISTURBANCE,2011/11/11 11:11:1,UNDER INVESTIGATION

Disturbance Analyzer Meas Result

Pressing this key selects Meas Result as the data type to be exported. Pressing this key when it is already
selected brings up the Meas Result Menu, which allows you to select and edit the contents of Meas Result
to save.
Key Path
Remote Command
Example
Notes
1106
exports the Disturbance List to the file myDisturbanceList2.csv in the default path.
The default path is at "My Documents\EMI\data\ DANalyzer \DisturbanceList\DisturbanceList_
xxx.csv"
If the save is initiated via SCPI, and the file already exists, the file will be overwritten.
Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an
instrument software upgrade.
Both single and double quotes are supported for any filename parameter over SCPI.
A.14.00
Save, Data
:MMEMory:STORe:RESults <filename>
:MMEM:STOR:RES "report1.html"
exports the measurement result to the file report1.html
The default path is at "My Documents\EMI\data\ DANalyzer \Results\report.html"