Thermo Scientific
b. Remove the filter from the filter bracket
Figure 39. Air filter location in the mass spectrometer with the front cover removed
Spring catches
2. Wash the air filter in a solution of soap and water.
3. Rinse the filter with tap water, and then allow it to air dry.
4. Reinstall the air filter and front cover.
5. Reconnect the plumbing tubing.
(Figure
39).
Air filter
Spring catches
TSQ Endura and TSQ Quantiva Hardware Manual
8
Maintenance
Maintaining the Air Filter
77