AFC-3000 Automated Fraction Collector

7.4 Installing the Kit for Normal-Phase LC

Description
Kit for normal-phase LC, including:
1.0 mm ID stainless steel drop former and grounding cable
Warning:
Avertissement: N'effectuez pas d'applications en phase normale sans vous être assuré
1. Switch the valve to Drain position, then shut down and unplug the fraction collector.
Set the pump flow to 0.
2. Carefully move the diverter valve assembly along the Y arm to the front.
3. Remove the current drop former. Then screw the SST drop former finger-tight into the
diverter valve.
4. Push the open clip end of the grounding cable from the normal-phase kit between the
diverter valve and the drop former (→ Fig. 25).
Page 64
Do not perform normal-phase applications without a proper
grounding protection of the drop former. Non-conductive, non-polar
normal-phase eluents, for example hexane, can lead to a build-up of
electrostatic charges on an ungrounded drop former. The electro-
static discharge can create a fire and/or explosion hazard.
que votre faiseur des gouttes est proprement isolée et mise à la terre.
Les solvants de la phase normale, non-conducteurs et non-polaires,
par exemple hexane, peuvent entraîner une accumulation de charges
électrostatiques sur le faiseur des gouttes en l'absence de mise à la
terre. Le décharge électrostatique peut représenter un risque
d'incendie et/ou d'explosion.
Fig. 25: Installing the normal-phase ground strip
Part No.
6702.0400
Operating Instructions