Gas Filter Interface Assembly Kit; Installation Procedure - Thermo Scientific Air-Jacketed Compact Flo-Thru 315 Operator's Manual And Parts List

Co2 incubators
Table of Contents

Advertisement

U
I
NPACKING AND
NSTALLATION
Warning
High concentrations of carbon diox-
ide product metabolic abnormalities,
disturbances of the central nervous
system and cardiac instability.
Unconsciousness may occur at con-
centrations above 10%.
6

Gas Filter Interface Assembly Kit

Models 315, 315-1, and 315DV require this assembly to be
used for both CO2 and air.
Kit includes:
• (4) Hose Clamps - #250-331-00
• (1) In-Line Filter - #525-034-00
• (1) 10 ft. (4.57 m) length of 1/4" (6.35 mm) ID
PVC tubing - #720-418-00

Installation Procedure

See next page for illustration.
1.
Turn the unit power switch to the OFF position.
2.
Connect tubing to tank gas outlet.
3.
Connect 1/4" (0.64 cm) ID tubing to hose barb on
one side of in-line filter and secure with hose
clamp; connect tubing to in-line filter hose barb on
its other side and secure with hose clamp.
4.
315 series incubators require this assembly to be
connected to both air and gas inlet fittings.
5.
Set the supply regulators at 4 psi.
6.
Check all of the connections for leaks.
7.
Turn the unit power switch to the ON position.