Installing The Oven Exhaust Kit - Thermo Scientific TRACE 1310 Hardware Manual

Gas chromatographs
Hide thumbs Also See for TRACE 1310:
Table of Contents

Advertisement

8
Installing Optional Kits

Installing the Oven Exhaust Kit

Installing the Oven Exhaust Kit
326
TRACE 1300 and TRACE 1310 Hardware Manual
The oven vents at the back of the GC discharge hot air up to 450 °C (842 °F) during cooling.
Installing the optional Oven Exhaust Kit (PN 19050760), the hot air from the oven vents can
be carried away to a fume hood or other exhausting devices
The kit includes two air ducts and two 3-m length extensible aluminium tubes.
See
Figure
325.
Figure 325. Air Duct and Extensible Aluminium Tube
Extensible Aluminium Tube
Air Duct
To install the oven exhaust kit
1. Put the GC in standby condition.
2. Cool the oven to room temperature.
3. Place each air duct on the oven vent inserting the hooks of the air duct into the relevant
vertical slots provided on the GC back panel.
4. Place an end of each extensible aluminium tube on each air duct.
5. Connect the other end of the extensible aluminium tube to an exhausting device.
6. Set the normal GC working conditions.
Extensible Aluminium Tube
Air Duct
Slot
.