Installing a sample compartment extension kit

Time needed: 10 minutes or less
Tools needed: 4-in1 tool (included with your instrument)
A 5/32 inch hex wrench
Note: To order parts, contact us.
1.—Remove any Smart Accessory, and then disconnect any cables that are connected at the
back of the sample compartment.
2.—Remove the stabilizer arm located on the top of the instrument at the back edge of the
sample compartment.
Remove the cap on top of the stabilizer arm, and then use the 4-in1 tool with a Philips bit
installed to remove the stabilizer arm screw. Lift the stabilizer arm off the instrument and
save it, with its cap and screw, in a safe place in case you want to reinstall it later.
3.—Place the sample compartment extension on the front of the sample compartment.
Make sure the pins in the same compartment base fit into the holes in the sample
compartment extension.
4.—Place the screws that came with the sample compartment extension kit into the holes in
the sample compartment extension, and then use the 4-in1 tool with a Philips bit installed
to tighten the screws. (The screws should be just finger-tight.)
5.—Install the sample compartment cover. You are now ready to install baseplate-mounted
accessories. Please see "Installing a sample" and "Adjusting sample height" for additional
information about using baseplate-mounted accessories.
Thermo Fisher Scientific
Using Your Nicolet iZ10 Module
15