Rap 5.19 J389- Holder Past Home Timeout; Rap 5.20 J390- Holder Rotate Horizontal Timeout - GBC eWire Service Manual

Table of Contents

Advertisement

RAP 5.19
J389- HOLDER PAST HOME TIMEOUT
J387 is triggered when it takes longer than specified time for the Holder to
reach S40 Holder near Element Feeder.
Follow the procedure in RAP 5.4
GBC eWire
RAP 5.20
J390- HOLDER ROTATE HORIZONTAL TIMEOUT
J390 is triggered if it takes longer than specified time for S35 Holder
Rotate horizontal sensor to get covered, to indicate Holder rotate is in
horizontal position.
1. Do GP 6.4 to check S35 Holder Rotate Horizontal sensor
This clears the fault
Yes- Return to normal operation;
No- Go to Step 2
2. Do GP 6.5 to check M14 Holder Rotate motor
This clears the fault
Yes- Return to normal operation;
No- Go to step 3
3. Check the presence and tightness of the two socket head screws on
the holder rotate crack
Screws are present and sufficiently tight
Yes- Go to Step 4;
No- Tighten the screw and resume normal operation
4. Check for the presence and tightness of the fasteners shown in the
below image at the Holder rotate pivot
The fasteners are sufficiently tight
Yes- Go to Step 5;
No- Tighten the fasteners and resume normal operation
01-26-2018
2-71
RAPS