5. Check if the set screws on all the six pulleys associated with the
Holder linear drive are tight
The set screws are tight
Yes- Go to Step 6;
No- Tighten the loose set screw(s) and resume operation
GBC eWire
6. Manually move the Holder linear mechanism and watch if the sheet
metal feature blocks/unblocks S40 Holder at Element feeder sensor.
The sensor is blocked/unblocked properly
Yes- Go to Step 7;
No- Escalate to second level
7. Do GP 6.5 to check M15 Holder linear drive motor
This clears the fault
Yes- Return to normal operation;
No- Go to step 8
8. Inspect the springs for the Holder gates (PL 5.21).
Holder gates springs are present and not damaged
Yes- Go to Step 9;
No- Replace the missing/damaged spring
9. Go to RAP 5.20, Step 5 and 6 to ensure the Holder rotate
components are in good order.
Holder rotate components are in good order
Yes- Go to Step 10;
No- Follow steps in RAP 5.20 to take the necessary corrective action
10. Do 6.3.7 FIRMWARE UPGRADE Procedure to Re-flash the firmware
for eWire.
This clears the fault
Yes- Return to normal operation;
No- Go to Step 11