8. If the detector is already installed, connect the reference electrode cable
and the cell cable.
9. To complete the cell plumbing, go on to
B.2

Installing Low-Pressure Valves in the DC

Three locations are available for installing low-pressure valves in the DC: on
the rear panel, in the lower compartment, and on the Dionex ICS-5000
Automation Manager (AM). To install low-pressure valves on the rear panel
or lower compartment, see the instructions in the next section. To install low-
pressure valves on the AM, see
B.2.1

Installing a Low-Pressure Valve on the DC Rear Panel

1. Insert the two tabs on the valve mounting plate into available slots on the
DC rear panel (see
Four low-pressure valves can be installed on the DC rear panel.
If an EO Regulator Kit (P/N 062582) is installed (see
two low-pressure valves can be installed on the EO bracket.
2. While holding the valve mounting plate firmly in place, tighten the
captive thumbscrew.
3. Plug the valve into one of the low-pressure valve connectors on the DC
rear panel (see
Doc. 065447-02 12/14
Section
B.4.
Figure
B-19) or EO regulator stand bracket.
Figure
B-19). Plumb the valve as required.
B • Installing DC Options
Section
4.1.13.