Start-up

Gas connection

Gas connection
Note Gas quality:
The gases must have one of the following qualities:
• Purity 99.5 % min
• medical gas quality.
CAUTION Overpressure!
The operating pressure of the gas applied to the device must not exceed 1 bar (14.5 psi). If the gas is
supplied at a higher pressure, the valves integral to the device may not close correctly and the gas
supply control may be impaired.
Set the gas supply to a range between 0.8 bar (11.6 psi) min. and 1.0 bar (14.5 psi) max. and make
sure that this pressure setting cannot be changed!

Installing gas pressure hoses

Figure 4-10. Installing gas pressure hoses

The gas supply from the gas supply system to the device is established using the supplied flexible gas
pressure hoses:
1. Connect the gas pressure hose to the sleeve of the gas supply system.
2. Remove the protective cap
3. Slide the hose clamp
to the sleeve
i160 / i250
S
4-10
TERI-CYCLE
(3/Figure
4-10) from the sterile filter.
(1/Figure
4-10) onto the gas pressure hose
(4/Figure
4-10) of the sterile filter
(2/Figure
(5/Figure
4-10).
4-10) and connect hose