Connecting the Servo Air for the Gas Sampling Valve

Connecting the Servo Air for the Gas Sampling
Valve
The connection of the servo air supply for the gas sampling valve is schematically
shown in Figure 4-6.
Use the following procedure to connect the servo air to the GC inlet manifold
when the GC is equipped with the gas sampling valve.
1. Make sure the air supply is turned off.
2. Connect the air line to the Servo air inlet with 1/8-in. Swagelok fittings.
NOTE
If the GC has been furnished with a 2x1 mm tube, use the ferrule for 2x1mm/1/8-in
provided in the start-up kit.
WARNING! Pressure of the servo air supply line should be set to a maximum of 500 kPa (72 psi).
Site Preparation and Installation Manual
Servo Air
Air
SERVO AIR
max 500 kPa
MAKE UP
H2
MAKE-UP
OUT
SAMPLE

Figure 4-6. Gas Sampling Valve Servo Air Connection

Plumbing the Gas Supply to the GC
!
1000 kPa/120 psi
CARRIER
Max for all inputs
H SENSOR
2
AIR
REF.
IN
SAMPLE
Chapter 4
63