TSQ 9000 Hardware Manual
To efficiently cool the TSQ 9000 instrument while it is being used, air must be allowed to
flow in through the front of the instrument and out the back. The instrument is equipped
with a fan that keeps the air flowing and filters that minimize the amount of dust that enters
the system. Dust can cause problems by insulating electronic components, which can lead to
overheating. To keep the instrument as cool as possible, we require that you keep all the covers
of the instrument attached while it is in use.
If the fan or filters in your TSQ 9000 instrument get damaged, use the illustrations below to
visually locate a component and then follow the process to replace it. If your filters are dirty,
but do not need to be replaced, see
Figure 416. Replaceable Fan and Filter Components of the TSQ 9000 System