Replacing Viper/nanoViper solvent lines

Thermo Scientific
6. Do one of the following:
• For a solvent-side check valve, reconnect the solvent inlet line.
• For a waste-side check valve, reconnect the waste line.
7. Mount the right side panel to the instrument housing. Then, with a #2 Phillips head
screwdriver, secure the panel by rotating the three captive screws a quarter-turn clockwise.
8. Run the Purge Solvent script for the pump associated with the replaced check valve, and
monitor the flow from the solvent bottle when the pump is filling and the flow to the
waste beaker when the pump is emptying (see
IMPORTANT The EASY-nLC 1200 requires nanoViper fitting systems with a specified
pressure range up to 1200 bar. Older versions are only specified up to 1000 bar. The new
1200 bar nanoViper fitting system consists of blue PEEK tubing, which distinguishes it
from the beige 1000 bar version, see
Figure 64. Visual difference between 1000 bar (top) and 1200 bar nanoViper fitting (bottom)
1200 bar version
For the EASY-nLC instrument, most of the plumbing connections are made with nanoViper
fittings
(Figure
65).
Note For more information about working with nanoViper fittings, see
nanoViper Fittings Quick Reference Guide"
Replacing Viper/nanoViper solvent lines
"Purge Solvent"
Figure
64.
on
page
273.
Troubleshooting & Maintenance Guide
5
Routine Maintenance
on
page
32).