Advanced Troubleshooting

Replacing Fans and Filters

Replacing Fans and Filters
354
TSQ 8000 Evo Hardware Manual
To efficiently cool the TSQ 8000 Evo instrument while it is being used, air must be allowed to
flow in through the front of the instrument and out the back. The instrument is equipped
with a fan that keeps the air flowing and filters that minimize the amount of dust that enters
the system. Dust can cause problems by insulating electronic components, which can lead to
overheating. To keep the instrument as cool as possible, we require that you keep all the covers
of the instrument attached while it is in use.
If the fan or filters in your TSQ 8000 Evo instrument get damaged, use the illustrations below
to visually locate a component and then follow the process to replace it. If your filters are
dirty, but do not need to be replaced, see
Figure 380. Replaceable Fan and Filter Components of the TSQ 8000 Evo System
Cleaning the
Filters.
Chassis Cooling
Fan
Front Wire
Mesh Filter
Turbo Cooling
Fan Filter