Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual page 144

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

4
Commands Quick Reference
Table 54 :TRIGger[:EDGE] Commands Summary
Command
:TRIGger[:EDGE]:COUPl
ing {AC | DC |
LFReject} (see
page
961)
:TRIGger[:EDGE]:LEVel
<level> [,<source>]
(see
page
962)
:TRIGger[:EDGE]:REJec
t {OFF | LFReject |
HFReject} (see
page
963)
:TRIGger[:EDGE]:SLOPe
<polarity> (see
page
964)
:TRIGger[:EDGE]:SOURc
e <source> (see
page
965)
144
Query
:TRIGger[:EDGE]:COUPl
ing? (see
page
961)
:TRIGger[:EDGE]:LEVel
? [<source>] (see
page
962)
:TRIGger[:EDGE]:REJec
t? (see
page
963)
:TRIGger[:EDGE]:SLOPe
? (see
page
964)
:TRIGger[:EDGE]:SOURc
e? (see
page
965)
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
Options and Query Returns
{AC | DC | LFReject}
For internal triggers, <level>
::= .75 x full-scale voltage from
center screen in NR3 format.
For external triggers, <level>
::= ±(external range setting) in
NR3 format.
For digital channels (MSO
models), <level> ::= ±8 V.
<source> ::= {CHANnel<n> |
EXTernal} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | EXTernal } for MSO
models
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
{OFF | LFReject | HFReject}
<polarity> ::= {POSitive |
NEGative | EITHer | ALTernate}
<source> ::= {CHANnel<n> |
EXTernal | LINE | WGEN} for the
DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | EXTernal | LINE |
WGEN} for the MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format

Advertisement

Table of Contents
loading