converter such as an integrator or other recording device. Refer to
and the documentation for the device for connection instructions.
The analog output voltages are updated with a data rate of 10 Hz (for the single-
channel Dionex VWD) or 50 Hz (for the four-channel Dionex VWD).
B.7.1

Installing the DAC Plug-In Module

1. Turn off the detector and disconnect the power cord from its source.
2. Use appropriate grounding protection to avoid electrostatic discharge,
which may result in damage to electronics components.
3. Remove the four screws indicated in
access panel on the detector's rear panel. Save the screws.
Remove
Screws (4)
Doc. 065377-01 1/12
Figure B-8. DAC Plug-In Module
Figure B-9. Removing the DAC Access Panel
B • Installation
Section B.7.2
Figure B-9
and remove the
91