Installing The Slide-In Module - Thermo Scientific thermoscientific Vanquish Pumps C Operating Manual

Table of Contents

Advertisement

7.16.3 Installing the Slide-In Module

Tools required
Preparations
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
NOTICE
Shipping the slide-in module improperly leads to damage to the device.
Always ship the slide-in module as described in this operating manual.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
Decontaminate all parts of the device that you want to return for
repair.
Fill in and sign the Health and Safety Form. Thermo Fisher Scientific
refuses to accept devices for repair if the Health and Safety Form is
missing, incompletely filled in, or unsigned.
CAUTION—Heavy Load, Bulky Device
The device is too heavy or bulky for one person alone to handle safely.
To avoid personal injury or damage to the device, observe the following
guidelines:
Physical handling of the device, including lifting or moving, requires
a team effort of two persons.
A team effort is in particular required when lifting the device into
the system stack or when removing it.
Screwdriver, Torx T20
1. Unpack the replacement device. Remove the slide-in module from
its enclosure as described in
(} page 191).
2. Verify that the device enclosure into which the slide-in module shall
be installed is clean. If required, clean the inner and outer surfaces
of the enclosure. See
(} page 126).
3. When installing the slide-in module to an enclosure in the system
stack, check that the enclosure is placed correctly in the stack.
7
Maintenance and Service
Removing the Slide-In Module
Cleaning or Decontaminating the Device
Page 193