Replacing The Flow Cell; Remove The Flow Cell - Thermo Scientific Dionex ICS Series Operator's Manual

Vwd variable wavelength detector
Hide thumbs Also See for Dionex ICS Series:
Table of Contents

Advertisement

Dionex VWD Operator's Manual
2. Flush the flow cell with deionized water until the solvent exiting the cell is
neutral (pH 7).
5.6

Replacing the Flow Cell

5.6.1

Remove the Flow Cell

When the flow cell is removed, light (at the wavelength set for the
detector) emits from the opening to the left of the flow cell. To avoid
possible damage to the eyes, turn off the lamps, turn off the power to
the detector, wear UV goggles, or set the wavelength to 600 nm or
above.
1. Unclip the cell inlet and outlet line unions (see
Cell Inlet and
Outlet Line
Unions
2. While holding the inlet union in place, disconnect the cell inlet line
from the union.
3. Repeat
4. Loosen the two flow cell screws.
5. Pull the flow cell out of the detector.
72
Figure 5-7. Removing the Flow Cell
NOTE Be careful not to twist, kink, or put other stress
on the factory-installed lines and connectors to
the flow cell
IN
Step 2
to disconnect the cell outlet line.
Figure
Flow Cell
Screws
and
OUT
ports.
5-7).