Wall Mounting; Quick Disconnect Fittings; Bench Mounting - Thermo Scientific 1267 Series Operating Manual

Reverse osmosis system
Table of Contents

Advertisement

Quick Disconnect

Bench Mounting

Wall Mounting

Thermo Scientific
These fittings are found on the prefilter, carbon and MPS cartridge, as well
as the water inlet. See Figure 4-4.
Fittings
To insert the cartridges or inlet tubing:
1. Press on the metal thumbpad on the unit to ensure the fitting is open.
2. Insert male adapter into fitting until you hear a click. Gently pull on
the cartridge or tubing to ensure it is secure.
3. To remove cartridges or tubing, press the metal thumbpad to release
fitting and pull adapter out.
PUSH DOWN ON RETAINER
TO RELEASE CONNECTOR.
Figure 4-4. Quick Disconnect
1. Place RO system on a bench top that is accessible to water,
atmospheric drain and electricity, and is convenient to your work area,
noting clearance requirements.
Install the RO system on a wall in a convenient location that is accessible
to water, atmospheric drain and electricity.
Caution Wall composition, condition and construction, as well as fastener
type, must be considered when mounting this unit. The mounting surface
and fasteners selected must be capable of supporting a minimum of 68 kg
(150 lbs.) Inadequate support and/or fasteners may result in damage to
mounting surface and/or equipment. If you are unsure of mounting
surface composition, condition and construction or correct fasteners,
consult your building maintenance group or contractor.
RETAINER
CONNECTOR
Section 4
Unpacking and Installation
s
Reverse Osmosis System
4-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

D126717157

Table of Contents