3
Components Installation

Installing the Valves

Installing the Valves

Installing Valves and Fittings into the Main Oven

Installing the Needle Valves and the Sample Inlet Fittings into the Main Oven
34
TRACE 1310 Auxiliary Oven Instruction Manual
This section provides instruction for installing and configuring the valves in the TRACE 1310
Auxiliary Oven.
I
Note Before starting, cool the TRACE 1310 Auxiliary Oven to 50 °C or cooler.
Do not turn the GC off.
The heated section of the TRACE 1310 Auxiliary Oven accepts standard VICI® gas actuated
valves with 3-inch standoff. The unheated section will accommodate VICI® gas actuated
valves without standoff.
CAUTION Be careful not to mix actuators with valves that have different numbers of
ports, as the actuator movement for each valve can be different.
Related topics:
"Installing Valves and Fittings into the Main Oven"
"Installing Valves and Fittings in the Unheated Compartment"
"Installing an Unheated Valve for Liquid Sample into the Unheated Compartment"
page 63
This section provides the instruction to install valves and sample inlet fittings into the main
oven.
Related topics:
"Installing the Needle Valves and the Sample Inlet Fittings into the Main Oven"
page 34
"Installing the Rotary Valves into the Main Oven"
"Installing the Diaphragm Valves in The Main Oven"
To install a needle valve and the sample inlet fitting into the main oven
Note The following example considers the installation of a needle valve into the seat 1
and the sample inlet fittings into the seat 5.
Note The sample inlet fittings (PN 19071014) are capable of permitting the introduction
of the sample up to a temperature of 100 °C. If you need to heat the sample inlet at higher
temperature than 100 °C before entering the main oven, the kit PN 19071022 is
required. See
"Bulkhead Connectors for High Temperature"
on
page 34
on
page 49
on
page 40
on
page 46
on
page
38.