Start-Up; Initial Operation; Installing Unit And Accessories - Thermo Scientific Safe 2020 Operating Instructions Manual

Safety cabinets
Table of Contents

Advertisement

5.

Start-up

5.1

Initial operation

Prior to initial operation, the safety cabinet must be subjected
to an installation test. Correct assembly and installation
performed by the operator are essential for good start-up.
Model Maxisafe 2020:
Upon the initial connection to the power supply system, the
device control of version Maxisafe 2020 starts an automatic ca-
libration routine to determine the parameters for the safety sys-
tem of the device.
NOTE – Calibration routine
The calibration routine is no substitite for the
installation check performed by the service
personnel.
5.2

Installing unit and accessories

Model Safe 2020:
Device without stand:
Place the device without stand onto a sufficiently stable
substructure so that the weight of the device frame does
not rest upon the floorpan.
Device with stand:
To assemble the (optional) stand and to install the device frame
to the stand:
1. Fig. 10a: Slide the two crossmembers [2] onto the retaining
angles [3] of the sidemembers [1]. Secure the crossmembers
to the two sidemembers using the screws [4].
2. To attach the device frame [5] to the stand [7], install four
Allen screws [6] loosely into the corresponding threaded
holes at the underside of the device.
3. Place the safety cabinet [1] onto the stand so that the
Allen screws [6] pass through the corresponding holes [8]
of the retaining tabs [10].
4. Push the device frame [5] in the grooves [9] of the retaining
tabs [10] all the way to the stop.
5. Tighten the four Allen screws [6].
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
Fig. 10a
Stand installation model
Safe 2020
25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Maxisafe 2020

Table of Contents