Installing The External Contact Interface Board - Thermo Scientific Agilent 1100 Series Getting Connected Manual

For lc devices
Table of Contents

Advertisement

2
Connecting the Hardware

Installing the External Contact Interface Board

Installing the External Contact Interface Board
8
Agilent 1100 Series Getting Connected Guide for LC Devices
If your Agilent 1100 Series autosampler is not already equipped with an external contact
interface board, install one as described below.
To install the external contact interface board
1. If you have not already done so, turn off the autosampler and unplug it from line power.
CAUTION To prevent personal injury caused by an electric shock, always turn off the
Agilent module and unplug its power supply cord from the electrical outlet before
removing the cover.
2. To prevent damage to the interface board, ensure that you are using ESD protection. For
more information on preventing damage caused by an electric discharge, refer to the
reference manual for the Agilent module.
CAUTION To prevent damage to an instrument, always use ESD protection when
handling electronic boards and components.
3. To remove the cover plate from the slot where you plan to install the external contact
interface board, loosen the two screws that fasten the plate to the chassis of the
autosampler.
4. Insert the external contact interface board into the slot (see
contact interface board part number). Tighten the two screws to fasten the board to the
chassis of the autosampler.
Figure 6
shows an installed external contact interface board.
Figure 6.
External contact interface board (BCD board) installed in an Agilent autosampler
REMOTE RS232
Screws that secure the cover plate
VIAL- # -OUTPUT
RELAY CONTACTS
CAN
CAN
Table 3
for the external
1
HP-18
CONFIG
8
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents