Prior to the initial operation, the biological safety cabinet must
be subjected to an installation test. Correct assembly and instal-
lation performed by the operator are essential for good start-up.
Upon the initial connection to the power supply system, the
device control starts an automatic calibration routine to deter-
mine the parameters for the safety system of the device.
NOTE – Calibration routine!
The calibration routine is no substitite for the
installation check performed by the service
personnel.
5.2

Installing the device and accessories

Device without stand:
Place the device without stand onto a sufficiently stable
substructure so that the weight of the device frame does
not rest upon the floorpan.
Remove the protective foil from the floorpan.
Device with stand:
Assemble the stand (accessory) and install the device frame
onto the stand:
1. Fig. 10: Slide the two crossmembers [2] onto the retaining
angles [3] of the side panels [1], then secure the crossmem-
bers to the side panels using the screws with wedge lock
washers [4].
2. To install the device frame [5] to the stand [7], insert four
Allen screws with wedge lock washers [6] loosely into the
threaded holes at the bottom of the device.
3. Place the biological safety cabinet onto the stand so that the
Allen screws with wedge lock washers [6] are routed through
the holes [8] of the retaining tabs [10].
4. Slide the device frame [5] into the retaining tab grooves [9]
all the way to the stop.
5. Tighten the four Allen screws with wedge lock washers [6].
Operating Instructions Biological Safety Cabinet | MSC-Advantage
Fig. 10
Stand installation
25