Calogix Setup

Programming the Controllers and Setting Parameters

Programming the
Controllers and
Setting Parameters
D-6
Omni FTIR Multi Gas CEMS Instruction Manual
5. The main Calogix control window will open. The window is a visual
representation of the hardware with readings and setpoints overlaid.
The following section details the procedure for setting up temperature
values and parameters in the CALogix controllers. These settings are pre-
configured into the system at the factory but may need to be set if a module
is replaced.
CAUTION The following procedures must be followed in the correct order.
The ramp program must be saved and the program mode must be set to
'On' before the controllers are set to run. If the controllers are set to run
with the ramp program not enabled, heated line and sample cell
temperatures could rise too quickly and damage those components.
Thermo Fisher Scientific