Getting Started

1. Verify that the detector gas control module installed on your GC is
2. The connections of the detector gases must be performed as described in the
For further details, refer to Chapters 16 and Chapter 19 of the TRACE GC Ultra
Operating Manual.

Transformer for NPD

To supply NPD, the dedicated transformer provided must be installed on the
Power Control Module of the GC.
WARNING! This operation must be carried out by authorized Thermo Fisher Scientific technical
personnel.

Getting Started

This paragraph contains the instruction to install the NPD on your GC.
To properly install the NPD upgrade kit, the following sequential operations must
be performed:
Installation Guide
compatible with the NPD gas requirements.
following table:
Installed
Connect Hydrogen
Module
AD
Gas 2
Remove the top, right side and rear panels of the GC.
Install the detector control card into the GC electronic compartment (control
unit).
Remove the GC Power Control Module
Mount the transformer for NPD.
Mount the NPD on the GC.
Remount GC panels.
Restart the GC.
Connect Air
to
Nitrogen Phosphorus Detector
Connect Make-up
to
Gas 1
Upgrade Kit
Gas to
Gas 3
7