Procedures For Connecting Two Or More Irs - FujiFilm CR-IR341 Service Manual

Fuji computed radiography
Table of Contents

Advertisement

3.7

Procedures for Connecting Two or More IRs

Procedures for connecting two or more IRs to a single CL are described below.
(1) Reconnect the network cable to another IR.
CL
(2) Set the CL configuration. The settings to be made are as follows.
• IR host name, IP address
• Application Entity name, service class, etc.
• Information on FINP for exchange of ID information
" Setting for the Destination (IR)" and "
in "3.1 Setting the CL Configuration Regarding IR Connection (Setting in CL's
Service Utility)
(3) For the built-in device, perform selector setting.
"3.2 Setting the Selector [Only When the Built-In Device Is Connected]"
(4) Set the IR configuration. The settings to be made are as follows.
• EQUIP file setting
• DEVICE file setting
• HOSTS file setting
• Base on DICOM file setting
"3.5 Setting the IR Configuration Regarding CL Connection (Setting in M-Utility)
(5) Check image transfer.
"3.6 Checking Image Transfer"
(6) If there is a third IR, repeat steps (1) through (5).
That's all there is to Step 2.
Perform image parameter transfer setting described in "3.8 Image Processing Parameter
Transfer Setting", as needed.
009-064-01
08.30.2001 FM3145
Setup completed
IR
Reconnect
No setup
IR
CR-IR341P/342P/343P/344P/345P/347P/348RU Service Manual
CL
IR
IR
00000054.EPS
Setting for ID Information Exchange"