Thermo Scientific
Operation
Preparing the environmental chamber
The environmental chamber must not be released for operation before all major start-up
activities have been completed.
Device Check
Prior to starting operation, the following environmental chamber components must be
checked for their correct function:
The door seal in the front frame must not be damaged.
The glass door must not be damaged.
The shelving components must be installed safely.
Disinfecting the work space.
Preliminary Cleaning and Disinfecting
Disinfect all interior surfaces with a general-use laboratory disinfectant, such as
quaternary ammonium, to remove any residues which may remain from production of the
environmental chamber. Rinse thoroughly with sterile distilled water, then 70% alcohol.
Dry with a sterile cloth as needed.
Disinfect the shelf channels and shelves, then rinse with distilled water before installing.
Before using any cleaning or decontamination method except those
recommended by the manufacturer, users should check with the manufacturer
that the proposed method will not damage the equipment.
Operation |
Chapter 6
Model 3962 | 6-1