RS-232 Communications
(continued)
Remote Alarm Contacts
and Analog Output
Thermo Scientific
The data transfer sequence is transmitted in the following format. X refers
to numerical temperature data.
(NUL) (-) XXX (SP) C (SP) (Error Message) (SP) (LF) (CR) (EOT) (SP)
In the event of a CNTRLFAIL, Er07, or the control probe is out of range
error, the numerical temperature data (XXX) in the transmission would be
replaced by T_ERR.
If no alarm condition exists, spaces will be sent. A total of 20 characters
will be sent.
SP - Space
CR - Carriage return
NUL - Null character (00)
If an alarm condition does exist, "Error Message" in the protocol will be
replaced by the following:
UNDERTEMP (temperature above the low alarm setpoint)
OVERTEMP (temperature below the high alarm setpoint)
PWRFAIL (AC power failure)
CNTRLFAIL (Control probe failure)
Er07 (micro failure)
* HSHX FAIL (Heat exchanger failure)
HOT COND (Hot condenser)
Note The RS-232 is not compatible with Model 1535 Monitor/Alarm
s
System.
* -86C units only
Revco CxF freezers have remote
alarm contacts and analog output.
See Figures 1-3 and 1-4 for the
location of the remote alarm
contacts. The remote alarm
connector is located in the parts bag
provided with the manual. It must
be installed if connecting the freezer
to an alarm system. After installing
the wiring from the alarm system to the connector, install the connector to
the freezer microboard and secure with the two screws provided. The
remote alarm provides a NO (normally open) output, a NC (normally
closed) output and COM (common). The contacts will trip on a power
outage, high temperature alarm or low temperature alarm. They will also
trip on high stage (-86C units only), control probe and microboard
failures. Figure 1-11 shows the remote contacts in alarm state.
LF - Line feed
EOT - End of text (4)
Figure 1-11. Remote Alarm Contacts
Section 1
Installation and Start-up
Revco CxF
1-9