Establishing Communications; Analog Output Installation - Thermo Scientific 5030 Instruction Manual

Synchronized hybrid ambient real-time particulate monitor
Hide thumbs Also See for 5030:
Table of Contents

Advertisement

Installation

Establishing Communications

Establishing
Communications

Analog Output Installation

2-22
Model 5030 SHARP Monitor Instruction Manual
21. Connect the 9-pin ambient/heater temperature cable assembly from the
heated sample tube to the 9-pin connector labeled "sensors" on the rear
of the monitor.
22. Place the heater control unit on top of the Model 5030 with the green
power switch facing forward.
23. Connect the brass resistance-heater connectors to the upper-most
exposed copper tubing and the lower-most exposed stainless sample
tube.
24. Connect one (1) blue/gray heater control cable to each brass resistance
heater connector. Wrap exposed stainless sample tube with insulation
provided.
25. Plug the heater control unit amphenol-connector into the rear of the
Model 5030 labeled "heater."
26. Check connections for proper installation.
27. Set monitor power switch to the ON position.
WARNING The Model 5030 Monitor and pump are supplied with
three-wire grounding cords. Under no circumstances should this grounding
system be defeated.
Unless specified upon order, the Model 5030 comes equipped with a default
analog output signal of 0-20 milliamps (mA).
This setting can simply be adjusted to a 4-20 mA setting via the user screen.
In order to configure the Model 5030 for an analog output of 0-10 volts
(V), jumpers on the main circuit board must be repositioned.
WARNING If the output signal jumpers are to be changed, TURN THE
POWER TO THE INSTRUMENT OFF FIRST AND UNPLUG FROM
THE POWER SUPPLY!