Chapter 2
|
Setup
Figure 2.2. TruFlow port on the controller.
Thermo Scientific
Figure 2.1. TruFlow gas manifold.
The gas supply to the TruFlow should be regulated to between 2 and
2.4 bar (30 and 35 psi) pressure. Since the TruFlow has no regulators,
the regulators must be fitted at the supply tanks.
Note: Regulators should be installed as close to the TruFlow as
possible as to avoid any flow problems.

2.4.1 Connecting the TruFlow

1. Ensure the TruFlow is placed in a secure location where it cannot tip
or be knocked over. Note: The TruFlow must be placed on the back
of the G3Lite cart.
2. Use the respective TruFlow communication cable to connect the
controller to the TruFlow. Refer to Figure 2.2 and Figure 2.3.
HyPerforma G3Lite Bioreactor Controller User's Guide
Figure 2.3. TruFlow communication ports.
Connect the
RJ45 connection
into the Ethernet
port on the
TruFlow
|
22