routine operations

Cleaning Baskets and Replacing Reagents

3-6 Cleaning Baskets
and replacing
reagents
3-6-1 running a cleaning
74
User Guide STP420D/STP420ES
1. Hang all the empty holders that were used in the embedding cycle
cycle
in the holder basket located in the rotation chamber or place the
unused holder in the biopsy chamber.
2. Close the lid and press the Clean quick start button on the Standby
screen. The 'Select chamber' dialog appears.
A "quick clean" of recently used chamber(s) performed by
default.
3. Press OK to confirm.
4. The text box shows the estimated finish time. Press Yes to confirm.
Note!
After 50 cycles of quick-cleaning you must perform an extended cleaning
cycle ("standard cleaning"). A reminder will appear on the screen.