Cleaning or Replacing the FID Collecting Electrode
Thermo Scientific
To clean or replace the collecting electrode
1. Put the GC in standby condition.
2. Turn the flame off. The fuel gases, hydrogen and air, are automatically closed.
3. Cool the detector to room temperature.
4. Turn the make-up gas off.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
5. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
6. If external modules are present, unplug the power cable from the AC Input connector of
each module, and from the wall outlet.
7. Open the module flap cover.
8. Disconnect the signal, glow-plug, and polarizing cables from their contacts on the cell.
Figure 117. FID Cables
4
Performing Detectors Routine Maintenance
Maintaining a Flame Ionization Detector (FID)
Signal Cable Straight Plug Crimp
Connector
Socket Set Screw
Glow-plug
Polarizing Cable Straight Jack Connector
TRACE 1300 and TRACE 1310 Hardware Manual
141