Replace a Broken Liner

OPERATING SEQUENCE
Replace a Broken Liner
Materials needed
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
1. While in stand-by condition, press OVEN to
access the oven control table.
and press OFF.
2. Scroll to
Temp
3. Press RIGHT INLET.
and press OFF.
4. Scroll to
Temp
Maintenance and Troubleshooting Manual
Screwdriver (P/N 205 026 03)
Liner removing tool (P/N 398 014 04)
Glass liner (see Table 9-2, on page 145)
Tweezers
OVEN
Temp
Initial time
Ramp 1
RIGHT INLET (PTV)
Temp
Pressure
Mode
Chapter 9
Maintaining a PTV Injector
.
50
50 <
1
10.0
50
50 <
100
100
split
157