Replacement Work
Replacing the Fuses
The two power fuses are located above the power switch [a] on the rear side of the microtome.
•
To replace the two fuses, turn off the power switch [a] of the instru- ment and unplug it.
•
Then open the cover [b] using a flat screwdriver.
•
Pull out the fuse holders and replace the fuses with new ones.
•
Put back the fuse holder completly and close the cover [b].
Rating of Fuses
For power requirements 100 – 240 V, 2 fuses T1,6AH, slow-blow.
56
Thermo Scientific Rotary Microtome HM 340E