Installing Sidewall Adapters - Thermo Scientific Nicolet iS10 Using Manual

Hide thumbs Also See for Nicolet iS10:
Table of Contents

Advertisement

5.—Use the 4-in-1 tool, with the small Philips bit installed, to tighten the screws that hold the
source in place. (Do not overtighten the screws.)
6.—Turn on your instrument and start OMNIC.
If you have installed a different type of source than you were previously using, choose
Experiment Setup in the Collect menu, and then click the Bench tab in the Experiment
Setup dialog box. Configure the Bench tab for the new source. (See the electronic
software documentation for more information.)
Use the Advanced Diagnostics command in the Collect menu to verify that the new source
is working properly. (If it does not work properly, contact us for assistance.)
7.—Make sure the system has been on for at least 15 minutes (one hour for best results), and
then align your instrument.

Installing sidewall adapters

Sidewall adapters are used to provide a tight purge seal for gas cells and other purged
accessories. To install sidewall adapters, press the adapter ring toward the sidewall to
compress the purge gasket, and then slide the adapter down into the clips on the side of the
transmission sample compartment.
Note: To remove a sidewall adapter, press the adapter ring toward the sidewall to compress
the purge gasket and then slide the adapter up out of the clips on the side of the sample
compartment.
Thermo Fisher Scientific
Using Your Nicolet iS10 Spectrometer
23