Upgrade Kit
Nitrogen Phosphorus Detector
Configuration
This paragraph contains the instructions to configure your GC to operate with
NPD. The following operations must be performed:
•
•
OPERATING SEQUENCE
Configuring Hardware, Detector and Make-up Gas
Hardware Configuration
This operation must be performed to verify if the GC has recognized the presence
of the transformer previously installed.
1. Press INFO/DIAG twice to open
2. Scroll to
3. Address the hardware configuration menu. Scroll down until the type of
4. If not, verify the transformer connections.
Detector and Make-up Gas Configuration
You configure the detector and make-up gas in the
Chapter 15 of the TRACE GC Ultra Operating manual.
1. Press CONFIG then scroll to
24
Hardware Configuration
Detector and Make-up Gas Configuration
Hardware Config
transformer installed in the system is displayed.
depending on the location of the detector to configure. Press ENTER to open
the detector gas menu.
menu.
Diagnostics
and press ENTER.
CONFIGURE
or
Left detector
Configuration
menu. Refer to
Right detector
Installation Guide