Performing A System Flush; Setting Timed Dispensing - Thermo Scientific Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List

Uv/uf ultrapure water system
Table of Contents

Advertisement

Setting Timed
Dispensing
Performing a System
Thermo Scientific
1. From the normal recirculating mode, when the display is showing
purity, press the UP or DOWN arrow until the display reads, "Auto
Dispense Menu?"
2. Press ENTER.
3. Display will read, "Dispense Method." Press the UP or DOWN arrow
to select TIME.
4. Press ENTER. Display will read, "Time Setting: xx Minutes."
5. Press the UP or DOWN arrows to set the desired time value. Press
ENTER.
UP arrow: +1 up to 5 min., then increments by 5 minutes
DOWN arrow: Decrement by 1 minute
Maximum: 40 minutes
Minimum: 1 minute
1. From the normal recirculating mode, when the display is showing
purity, press the UP or DOWN arrow until the display reads, "System
Flush
Flush?" Press ENTER.
2. Press the UP or DOWN arrow to select "Short Flush?" or "Extended
Flush?" Press ENTER.
3. Display will read, "System Flushing: (xxx) Seconds or (xxx) Minutes
Remaining."
Note An automatic 1-minute flush will occur once every 12 hours. If a
unit is (Idle) or in STANDBY for over 12 hours, an automatic system
flush will occur immediately after the START/STOP switch is pressed.
Note Press the STOP/STOP switch to cancel the flush.