High Temperature Cutoff Switch; Demineralizer Cartridge Changing - Thermo Scientific 678 Series Operating Manual And Parts List

Table of Contents

Advertisement

Demineralizer
Cartridge Changing
High Temperature
Cutoff Switch
Thermo Scientific
When one of the disposable demineralizer cartridges expires, the red
"CHARGE CARTRIDGE" light will come on and the water still will shut
down. Replace the cartridge as follows:
1. Turn off still and water supply.
2. Open the cartridge drain pinch clamp located behind the cartridges.
Knobs on the cartridge holders may be loosened to speed up draining.
3. Remove and discard left hand cartridge.
4. Remove right hand cartridge and install in left hand holder. Tighten
cartridge holder knob.
5. Place new cartridge upright in the right hand holder and tighten
cartridge holder knob. Close drain pinch clamps.
6. Proceed through Steps 4 - 10 in the operating instructions. Flow
meters should not need adjustment.
Description
High Purity Cartridge
High Capacity Cartridge
Your MEGA-PURE 12A Water Still is protected against overheating by
two thermal switches, located in the black "V" shaped bracket at the left
hand side of the boiler, that will shut off both the water supply and power
to the heaters when activated. Always allow the unit to cool completely (5
- 15 minutes) before resuming operation. Cool water on hot glass could
cause boiler damage. When unit has cooled, press "WATER" and
OPERATE" switches to restore normal operation. Check boiler
occasionally for proper water level.
Caution Always allow the unit to cool completely before resuming
operation. Cool water on hot glass could cause boiler damage.
Deionizer Cartridges
Use
Used for high purity water
Removes ionized solids for extended capacity
Section 6
Operation
Part #
D4000377
D400499
s
12A Water Still
6-3