Environmental Requirements - GE Carestation 620 User's Reference Manual

Table of Contents

Advertisement

Environmental requirements

Temperature
Humidity
Altitude (with airway
module)
Altitude (without airway
module)
2076152-001
.
Operation
10 to 40°C
Oxygen cell 10 to 40°C
15 to 95% RH, non-condensing
500 to 800 mmHg
(3565 to -440 meters)
475 to 800 mmHg
(4000 to -440 meters)
Note
Do not operate the machine, or apply mains power, when the system
is outside the specified operational temperature range and specified
operational humidity. If the device is colder or warmer than the
specified operating temperature, or if condensed water is visible or
suspected, let the temperature stabilize before use.
11 Specifications and theory of operation
Storage and Transport
-25 to 60°C
Oxygen cell storage is -15 to 50°C, 10 to 95% RH
LCD display storage is -20 to 60°C
15 to 95% RH, non-condensing
425 to 800 mmHg
(4880 to -440 meters)
425 to 800 mmHg
(4880 to -440 meters)
11-19