Overview; System Specifications; Classification Type; Environmental Requirements - GE Brivo OEC 715 Operator's Manual

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

Technical Reference

12.1. Overview

The policy of GE Healthcare is one of continual product development and improvement.
For this reason, GE Healthcare reserves the right to change the operating characteristics and specifications
of newer products at any time, without prior notice, and without incurring any obligation relating to
previously manufactured items.
The specifications listed are limited to general performance and physical data. Specifications of optional
equipment provided by other manufacturers are given in the applicable manuals provided with those
options.

12.2. System Specifications

12.2.1. Classification Type

Class I equipment (as defined by IEC60601-1)
Type B protection against electric shock (IEC60601-1)
Ordinary protection against ingress of water
Degree of Protection:
System: IPX0
Footswitch: IPX8
Operation Mode:
Continuous operation
Loading Mode:
Intermittent loading
All parts of the equipment, including accessories and connecting cables, can be disinfected by wiping
them with a cloth dampened with disinfectant, such as 75% alcohol.

12.2.2. Environmental Requirements

Operating Conditions:
Temperature:
Humidity:
Atmospheric Pressure:
Transportation and Storage:
Temperature:
Humidity:
Atmospheric Pressure:
Page 12-2
+10°C~+40°C (+50F~+104F)
20%~80% Relative, non-condensing
700hPa~1060hPa
-20°C~+55°C (-4F~+131F)
10%~95% Relative, non-condensing
500hPa~1060hPa