General Console Requirements; Console Environmental Requirements - GE LOGIQ 5 Service Manual

Hide thumbs Also See for LOGIQ 5:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300000, R
IRECTION
EVISION
Section 2-2

General Console Requirements

2-2-1

Console Environmental Requirements

Table 2-9
Temperature
Humidity
Pressure
Table 2-10
Power Source
Current Rating
Radiation Shielding
Heat Dissipation
Floor Landing
Floor Condition
2-2-1-1
Cooling
The cooling requirement for the LOGIQ™ 5 is 3500 BTU/hr. This figure does not include cooling needed
for lights, people, or other equipment in the room. Each person in the room places an additional 300
BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and diameters can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interface.
2 - 2
1
Environmental Requirements for LOGIQ™ 5 Scanners
Operational
o
10 - 30
C
o
50 - 104
F
30 - 85%
non-condensing
700 - 1060hPa
Environmental Requirements for an Ultrasound Room
Item
Refer to Table 2-11 on page 2-3.
15A (120V-100V); 8A (220-240V) CIRCUIT BREAKER
NONE REQUIRED for ULTRASOUND ENERGY
Temperature
20-26 DEG. C (68-79 DEG F) for PATIENT COMFORT
Humidity
50% to 70% for PATIENT COMFORT
2000 BTU/Hr.
Approximately 680 - 800 kg/m
Gradient: WITHIN 5 degrees
Weight
180 kg (397lbs) without Accessories
Section 2-2 - General Console Requirements
Storage
o
-10 - 60
C
-40 - 60
o
14 - 140
F
-40 - 140
30 - 90%
30 - 90%
non-condensing
non-condensing
700 - 1060hPa
700 - 1060hPa
Values
2
without Accessories
LOGIQ™ 5 S
ERVICE
Transport
o
C
o
F
M
ANUAL