General Logiq Totus Requirements; Logiq Totus Environmental Requirements - GE LOGIQ Totus Basic Service Manual

Table of Contents

Advertisement

Direction 5936427, Rev. 1
Section 2-2

General LOGIQ Totus Requirements

Contents in this section
2-2-1
2-2-2
2-2-3
2-2-4
2-2-5
2-2-1

LOGIQ Totus environmental requirements

If the LOGIQ Totus is very cold or hot
RISK OF DAMAGE TO EQUIPMENT
CAUTION
IF THE ULTRASOUND SYSTEM IS VERY COLD OR HOT, DO NOT TURN ON ITS POWER UNTIL
IT HAS HAD A CHANCE TO ACCLIMATE TO ITS OPERATING ENVIRONMENT.
When unpacking the LOGIQ Totus, allow the temperature to stabilize before powering up.
Refer to
temperatures.
Table 2-1
o
C
o
122
F
Hrs
Environmental specifications
The LOGIQ Totus should be operated, stored, and transported according to the listed environmental
requirements.
NOTE:
Probe environmental specifications may be different than the LOGIQ Totus.
Refer to
Table 2-2
Temperature
Humidity
Pressure
Heat Dissipation
2 - 2
GE HealthCare CONFIDENTIAL
LOGIQ Totus environmental requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Electrical requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
EMI limitations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Probe environmental requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Time and manpower requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
for guidelines on reaching operational temperatures from storage or transport
Table 2-1
LOGIQ Totus Acclimate Time
50
45
40
35
30
113
104
95
86
4
2
0
0
0
for environmental specifications.
Table 2-2
Environmental Specifications for the LOGIQ Totus
Operational
10° - 35°C
50° - 95°F
30 - 80% non-condensing
70 - 106kPa
3070 Btu/hour
Section 2-2 - General LOGIQ Totus Requirements
25
20
15
10
5
77
68
59
50
41
0
0
0
0
2
Storage
-10° - 50°C
14° - 122°F
10 - 90% non-condensing
70 - 106kPa
n/a
LOGIQ Totus Basic Service Manual
0
-5
-10
32
23
14
4
6
8
Transport (< 16 hrs.)
-10° - 50°C
14° - 122°F
10 - 90% non-condensing
70 - 106kPa
n/a