Line supply cable

ILLUSTRATION 2-2
LINE SUPPLY CABLE
3–5–1 Obtaining a line supply cable
3–6

Main circuit breaker

Note:
The line supply cable comprises of two supply wires and a ground cable (i.e. 3 x AWG 10 (5.32
2
mm
)) with the following actual/usable lengths:
- Total length = 7 m (23'),
- Usable length = 6.5 m (21'-4").
D
For customers in all countries except China, the optional line supply cable (S30331BC)
must be ordered from the Price Book so that it is supplied with the Senographe System.
D
For customers in China, the optional line supply cable from GEMS must not be ordered.
Instead Chinese customers must locally order an equivalent CCC-certified supply cable
from their local electrical supplier.
ONLY for GEMSE: the Generic EDM is associated with an external modem (Motorola
3265) mounted on the EDM cabinet.
For GEMS A and GEMS AM the Generic EDM incorporates a modem board inside the
The main circuit breaker is supplied by the customer and must be
EDM cabinet.
sized in accordance with local regulations.
Circuit breakers sizes for European market:
- From 380 V up to 415 V: circuit breaker: In = 15 A - magnetic I = 7 In ±20%,
- From 200 V up to 240 V: circuit breaker: In = 20 A - magnetic I = 7 In ±20%.
Circuit breakers size and supply conductors for US market: refer to Section 517–71(a)
and Section 517–73(a) (Item 1, 2) of the NEC–1993 (see below).
- The branch circuit used must be rated 30 A or less.