1.2.3 Innova Patient table
Innova Systems can be supplied with an Omega table, which includes a table side control with
Contour Filter.
1.2.4 X–Ray generator
The Innova System uses two Innova High Voltage Generator System with component parts as
follows according to the power requested (100 kW):
•
Two Jedi – C1 Frontal and C1 Lateral cabinets,
•
Two Ingrid H.V. tank mounted on the Ulysses X–Ray tube housing.
•
Two External recirculating chillers (mandatory for each tube). It is mandatory to place the
two Chillers in the Technical Room.
1.2.5 X–Ray head
The Innova System uses two Ulysses X–Ray tube housing including an X–Ray tube, a
collimator, Ingrid HV tank, oil/water exchanger, contour filter, etc.
54
Innova
IGS 620, Innova
TM
Illustration 2-3:
IGS 630 Preinstallation Manual
TM
Direction 5694389-1-1EN, Revision 2
1 System Components