Description of the X–Ray generator
The Innova 2000 System uses an Innova High Voltage System with component parts as follows
according to the power requested (100 kW):
D MPPU1 – Cabinet with X–Ray generator,
D MPPU2 – Cabinet with AC distribution panel,
D MPPU5 – Microgrid H.V. tank mounted on the Maxiray150 X – Ray tube housing.
2-4
Description of the X–Ray head
The Innova 2000 System uses a MAXIRAY 150 tube housing including an X–Ray tube, a SIEMENS
collimator, VMP micro grid HV tank, oil/water exchanger, contour filter, etc.
Note:
An external recirculating chiller is mandatory.
The Chiller elevation cannot exceed 3 meters (210 feet).
CAUTION
Depending on the location of the system, you may choose:
D either low noise chiller (if placed in an exam room),
D or a high ambient chiller (if placed in a technical room).