Instrument Messages
Advisory Messages
Message
Fixed LO freq should be
greater than RF Stop freq
Freq Scale Type=Log is not
available in Zero Span
Frequency Hopping enabled,
waiting for valid burst
Frequency menu has changed
to reflect frequency
context switch
Gate required for valid
results
High-pass and Low-pass
filters set to OFF
High-pass filter set to OFF
IF Fixed freq should be
greater than LO Stop freq
IF Fixed freq should be
greater than RF Stop freq
IF Start freq should be
greater than LO Fixed freq
IF Start freq should be
greater than RF Start freq
Input is internal
Instrument Messages
Description/Correction Information
The setup frequencies break the rules for a downconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The LO fixed freq should be greater than the RF freq's for an LSB or
DSB (for DSB measurements the setup uses LSB values) downconverter setup. Use
the graph icon on the DUT setup form to clarify the setup required.
Logarithmic scaling cannot be used for time domain sweeps (0 Hz span).
The demodulated burst type has not been found in the originally demodulated slot
location within the frame.
The frequency context parameter has been changed either by the user or the
system. The frequency menu will now contain the frequencies for the new context.
No action required.
Turning on any band pass filter will turn off high-pass and low-pass filters.
Turning on any band pass filter will turn off high-pass filters.
The setup frequencies break the rules for an upconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The IF fixed freq should be greater than the LO Stop freq for a USB
upconverter swept LO setup. Use the graph icon on the DUT setup form to clarify
the setup required.
The setup frequencies break the rules for an upconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The IF fixed freq should be greater than the RF Stop freq for an
upconverter swept LO setup. Use the graph icon on the DUT setup form to clarify
the setup required.
The setup frequencies break the rules for an upconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The IF start freq should be greater than the LO fixed freq for an USB
upconverter fixed LO setup. Use the graph icon on the DUT setup form to clarify the
setup required.
The setup frequencies break the rules for an upconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The IF start freq should be greater than the RF Start freq for an
upconverter fixed LO setup. Use the graph icon on the DUT setup form to clarify the
setup required.
The instrument's input is set to internal (the internal amplitude reference signal). So
any signals connected to the front/rear panel inputs cannot be measured.