Figure 6-5 Top View With Top Clamshell Removed - Keysight N1913B EPM Series Service Manual

Power meters
Table of Contents

Advertisement

Instructions
Key to
Figure
6-5:
1 Front panel assembly
2 Calibrator assembly
3 Sensor flex connection(s)
4 Calibrator cable connection
5 Rear panel assembly
6 Line module
7 Fan assembly
8 Analog recorder output connection(s)
9 VGA cable
10 USB cable
11 P700 carrier board
12 Trigger BNC
With reference to
Figure
– Lift and remove the cable clamp.
– Disconnect the cable attaching the fan assembly to the main
board.
– Lift and remove the fan assembly
– Disconnect the sensor RF connections from the main board.
– Disconnect the sensor flex connection(s) from the main
board.
– Disconnect the calibrator cable connection from the main
board.
– Disconnect the analog recorder output connection(s) from
the main board.
Keysight N1913B/N1914B EPM Series Power Meters Service Guide
6-5:
Visual
1
2
3
4
Figure 6-5
Top view with top clamshell
removed
Repair Guide
7
6
8
10
5
11
9
9
12
107
6

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1914b epm series

Table of Contents